![GeminiBio MXG102 Скачать руководство пользователя страница 89](http://html1.mh-extra.com/html/geminibio/mxg102/mxg102_user-manual_2207788089.webp)
Moxi GO II
™
User Guide
Page 84
Counts
Optimal fluorescence
Type S+ cassette
Counts
Optimal fluorescence
Measurement Time
1 x 10
4
– 1.75 x 10
6
cells/ml
< 5 x 10
5
cells/ml
1 x 10
4
– 7.5 x 105 cells/ml
< 3 x 10
5
cells/ml
~ 5-10 seconds
Software
Includes a (pre-loaded) instrument-only FW/SW
package that drives instrument performance and on-
instrument data analysis. Periodic software updates
are made available via
www.orflo.com
. These
updates are offered both to fix bugs as well as to
enable new functionality.
Offline analysis requires FCS v3.1 compatible
software package for viewing files (e.g., FlowJo X or
FCSExpress 7)
Physical Attributes
Benchtop
Dimensions (HxWxD)
Weight
Yes
5.8 in x 9.3 in x 8.7 in / 14.8 cm x 22.1 cm x 23.6 cm
10 lbs / 4.54 kg
Electrical Specifications
Internal Battery
AC Power Adapters (US, EU, and
UK types)
Rechargeable 3.7V, 7500 mAh lithium ion
Input: 100-240 VAC (50/60 Hz), 0.2 A
Output: 5 VDC, 1 A
Operating Environmental
Conditions
Temperature
Maximum Relative Humidity
Altitude
Applicable Pollution Degree
Degree of Ingress Protection
15-30
o
C (indoor use only)
20-80% (non-condensing)
2000 meters (maximum)
Pollution Degree 2
IPX10
Cassette Handling Storage
Cassettes will last indefinitely when handled/stored
properly (no cigarette smoke, no exposure to extreme
temperature cycles, general storage temperature of
15-30C). The cassettes can be handled/touched
anywhere on the cassette, with the exception that we
recommend minimizing contact in the black regions
towards the center of the cassette as those are the
optical regions.
Содержание MXG102
Страница 68: ...Moxi GO II User Guide Page 63...
Страница 75: ...Moxi GO II User Guide Page 70...