Moxi GO II
™
User Guide
Page 6
Part
Function
Power Adapter, US Style
Connects USB cable to a 110 VAC outlet
USB Cable
Connects instrument to Personal computer
(PC or Mac) or power adapter
Cassette Tray Insert
Extra tray insert can be user-installed in
cassette tray when adhesive build-up
interferes with insertion/removal of cassettes.
Getting Started
The Moxi GO II™ is shipped in a condition ready for initial use; though we recommend checking
for software updates and updating the Moxi GO II with the most current software version upon
receipt. To begin, plug the USB Cable into the instrument at one end (mini-USB). Plug the
opposite end of the USB Cable (standard USB) into the Power Adapter. Plug the Power
Adapter into an AC outlet that matches the rating of the Power Adapter.
CAUTION
Use only an ORFLO supplied USB Cable and Power Adapter. Use of
other products may result in inaccurate test results or damage to the
instrument, and will void warranty.
NOTES
Tests should only be performed with the instrument powered by an AC
source.
The device will power on and users can access files without attaching to a
power outlet but for data collection the device must be plugged into an
electrical outlet.
General Guidelines
The Moxi GO II is used with Type S+ or MF-M Cassettes. Refer to the Specifications section for
information on the operating size and concentration ranges for the system.
The sample volume for a test should be 60 - 75 µL.
Cells should be suspended in 0.9% salt solution (e.g. PBS or equivalent) to ensure proper
conductivity for unit operation and to ensure proper particle sizing. Water, hypotonic, or
hypertonic solutions are
not
acceptable diluents.