Moxi GO II
™
User Guide
Page 78
Small Particles
A smaller particle population might have a left tail
that extends into the system noise (below the
specified range of the cassette). Re-scaling the
x-axis (can only be done after a test, prior to
saving) can increase the binning resolution
around these populations, helping to better view
the peak.
Debris-filled sample or size-
overlapping cell samples
(polydisperse samples).
Overlapping size populations can obscure the
downward slopes of cell peaks. Often debris
populations can obscure the left side of a cell
peak.
Overly Long Test Times
Insufficient fluid (<60 µL)
Adjust pipette to ensure sufficient volume for the
test. For improperly calibrated pipettes, this
might require a setting >60 µL
Cells/beads too large
Make certain that the particle type is within the
specified size range of the cassettes being used
Highly/strongly aggregated cells
clogging filter and blocking flow
Try breaking apart cells using pipette trituration
and/or protease treatment.
Blue Screen – “Software
Exception”
System firmware has detected a
malfunction.
Please take a photo and send to
[email protected]. Reset the instrument
by pressing and holding the power button for at
least 5 seconds. When plugging the system into
a computer to access the data, the system may
report a crash log has been generated. Email
that file “OS_Syslog.txt” to
[email protected]
System shuts off during
alignment/test
System is unplugged
Plug the system in to the AC wall adapter (NOT A
COMPUTER USB PORT) and re-try. Make sure
the battery is fully charged. If the battery is not
fully charged, keep the system plugged into the
wall outlet, make sure it is powered on, and
charge it overnight
Error Messages
Cause
Corrective Action
Alignment Failed: Cassette
Aperture Not Found
Cassette error
Try the other side of the cassette or a new
cassette
Calibration: Please remove the
cassette
Cassette in tray on startup
Remove the cassette from the tray and tap "OK"
Error: The disk is full
Not enough space to save a test.
Delete files from the "Saved Tests". Selecting
“Delete All” and “Bitmaps Only” can be a method
to quickly clear space on your disk without
removing the data files. The Bitmap files are the
largest files on the disk so, if there are a lot of
them, this option will be particularly effective.
Invalid or empty filename
Filename field is blank
Enter a valid filename.
No Suitable Fit
No clear Gaussian population in a
Size Histogram
Revert to gating mode to obtain sizing and
counts.