Direction 5743854-1EN, Revision 1 Optima XR642/XR648 Pre-Installation
Chapter 4 - HVAC
Page 93
Chapter 4 HVAC
Note: Table information is only applicable for Optima XR648.
Section 1.0 Environmental Requirements
1.1
Relative Humidity and Temperature
Limits for rates of change:
In-Use Storage
< 10 degree C / hour < 20 degree C / hour
< 30% / hour < 30% / hour
STORAGE values only refer to equipment that is still in shipping containers. If the equipment is partially or completely
installed, refer to IN-USE values.
The operating environment is the environmental requirements for the normal use of the device during the expected
service life. It dose not apply to the short term manufacturing testing and the like.
Table 4-1 Environmental Requirements (Relative Humidity & Temperature)
Product or Component
RELATIVE HUMIDITY
(Non-Condensing)
Temperature
IN-USE
IN-USE
MIN
MAX
MIN.
MAX.
URP Detector
30%
60%
15° C (59° F)
35° C (95° F)
Wall Stand Std.& Man. & Ext.
30%
60%
15° C (59° F)
35° C (95° F)
Table (TBL)
30%
60%
15° C (59° F)
35° C (95° F)
OTS
30%
60%
15° C (59° F)
35° C (95° F)
System Cabinet(SKL1)
30%
60%
15° C (59° F)
35° C (95° F)
Maxiray 100-09 X-ray Tube
-
-
15° C (59° F)
35° C (95° F)
Radiographic Stretcher Table (optional)
30%
60%
15° C (59° F)
35° C (95° F)
Operator Console:
PC Tower
30%
60%
15° C (59° F)
35° C (95° F)
LCD Monitor
30%
60%
15° C (59° F)
35° C (95° F)