GE Healthcare
Direction 5743854-1EN, Revision 1 Optima XR642/XR648 Pre-Installation
Chapter 6 - Communications/Networking
Page 113
most popular networking connection to ease integration into your facility's network.
In the system room, this facility;
Do you segment your network using subnets?
Our equipment's IP addresses are:
2.5
System Parameters
The system default uses the following IP Addresses internally:
•
192.168.1.1 eth0
•
192.168.2.1 eth1
•
192.168.3.1 br0 (eth2/3)
•
192.168.1.30 detector IP before detector boot
•
192.168.1.50 detector IP after boot in table
•
192.168.2.45 detector IP after boot in WS
•
192.168.3.50 detector IP after boot in Tether/wireless
If the hospital network uses 192.168.x.x, there will be a conflict. If this conflict occurs, you must
contact your GE Service Representative to change the internal IP Addresses used by the system.
System
Host Name: _________________________
Network (IP) Address: _ _ _ . _ _ _ . _ _ _ . _ _ _
Subnet Mask: _ _ _ . _ _ _ . _ _ _ . _ _ _
Router IP: _ _ _ . _ _ _ . _ _ _ . _ _ _
Scheduled Station AE Title: _________________________
Has 100baseT installed
Has 10baseT installed
Has a different network installed
Will have 100baseT installed
Will have 10baseT installed
We don't have a network installed
Yes
No
Static
Acquired via DHCP
A combination of both methods
Pre-Install.book Page 113 Tuesday, January 30, 2018 2:36 PM