GE Healthcare
Direction 5743854-1EN, Revision 1 Optima XR642/XR648 Pre-Installation
Page 76
Section 3.0 - Room Layout
3.2
Clinical Access
Make sure that you plan the room with the following clinical access requirements:
• Provide easy access to the patient table. Stretchers and other mobile hospital equipment must
reach the table quickly.
• Table cannot be installed at 90 degrees to the ceiling stationary rails.
• When installing an extended-arm Wall Stand, the Wall Stand should be positioned directly in
the center of the 3-meter/4-meter bridge if full tube angling capabilities are to be achieved.
• The Wall Stand can be configured to allow grid insertion direction from either the left or the
right. Workflow and room dimensions should be considered.
3.3
Peripheral Equipment
Consult hospital personnel regarding additional space requirements for the following types of
hospital equipment:
• Storage Cabinets
• Sinks
• Oxygen Stations
• Monitoring Equipment
• Crash Cart
3.4
Dongle default location and adjusting range
Dongle ASM location which is critial for wireless signal quality, other components can be located anywhere cable length
allows and customer prefer for application. So this section just describe room layout requirements for Dongle ASM
installation.
3.4.1
Inside control room layout:
1.) If the most common used digital cassette application is area-1, select position1 as the default dongle installation
position; If it’s area2 use position2 as dongle default installation position
2.) Position1 is in the center from point A to B, that means it’s the center position from edge of control room wall to
exam room wall. The height requirement of dongle is 30cm lower than the ceiling. There is the possible adjusting
Pre-Install.book Page 76 Tuesday, January 30, 2018 2:36 PM