Section 4.0 - Product Storage and Handling Requirements
4.2
Atmospheric Pressure
Limits for rates of change:
STORAGE values only refer to equipment that is still in shipping containers. If the equipment is partially or
completely installed, refer to IN-USE values.
Table 1-3 Environmental Requirements - (Altitude & Atmospheric Pressure)
PRODUCT OR COMPONENT ALTITUDE
ATMOSPHERIC PRESSURE
Stationary Operating Environment
and Non-operating Environment
MIN
MAX
MIN
MAX
Total System Limits
-30 m (-98.43 ft)
3000 m(9,842.52 ft)
70 kPa
106 kPa
In-Use:
Storage:
<1.8 kPA / hour
<76 kPA / hour