GE Healthcare
Direction 5743854-1EN, Revision 1 Optima XR642/XR648 Pre-Installation
Page 94
Section 1.0 - Environmental Requirements
1.2
Altitude and Atmospheric Pressure
Limits for rates of change:
In-Use Storage
< 1.8 kPA / hour < 76 kPA / hour
STORAGE values only refer to equipment that is still in shipping containers. If the equipment is partially or completely
installed, refer to IN-USE values.
The operating environment is the environmental requirements for the normal use of the device during the expected
service life. It dose not apply to the short term manufacturing testing and the like.
1.3
Heat Output
The standby and in-use heat output of each system component are provided below.
1.4
EMC Requirement
Refer to the Direction 5743857-1EN Optima XR642/XR648 Specification Chapter 4 - EMC Compliance Statement
Table 4-2 Environmental Requirements - (Altitude & Atmospheric Pressure)
Product or Component ALTITUDE
ATMOSPHERIC PRESSURE
IN-USE
STORAGE
IN-USE
STORAGE
MIN
MAX
MIN
MAX
MIN
MAX
MIN
MAX
Total System Limits
-30 m
(-98.43 ft)
3000 m
(9843 ft)
-30 m
(-100 ft)
3000 m
(9843 ft)
70 kPa
106 kPa
70 kPa
106 kPa
Table 4-3 Heat Outputs by Component
PRODUCT OR COMPONENT
HEAT OUTPUT
STANDBY
IN-USE
WS Detector power
56 BTU/h
0.017 Kilowatt
56 BTU/h
0.017 Kilowatt
Manual / Standard / Extended Wall Stand
79 BTU/h
0.023 Kilowatt
321 BTU/h
0.094 Kilowatt
TABLE Detector power
56 BTU/h
0.017 Kilowatt
56 BTU/h
0.017 Kilowatt
Table
315 BTU/h
0.092 Kilowatt
2272 BTU/h
0.666 Kilowatt
OTS & Collimator
105 BTU/h
0.031 Kilowatt
105 BTU/h
0.031 Kilowatt
Tube Rotor
0
0
544 BTU/h
0.160 Kilowatt
System Cabinet
2437 BTU/h
0.714 Kilowatt
4869 BTU/h
1.427 Kilowatt
Z420 PC+ Monitor
601 BTU/h
0.176 Kilowatt
863 BTU/h
0.253 Kilowatt
TIB
6.75 BTU/h
0.002 Kilowatt
68 BTU/h
0.020 Kilowatt
UPS
31.61 BTU/h
0.009 Kilowatt
45.45 BTU/h
0.013 Kilowatt
Pre-Install.book Page 94 Tuesday, January 30, 2018 2:36 PM