GE Healthcare
Direction 5743854-1EN, Revision 1 Optima XR642/XR648 Pre-Installation
Chapter 2 - Equipment
Page 85
Illustration 2-42 : Manual WS only with 2M bridge
Item
Limit
Dimension
Recommend
Dimension
Description
mm
inch
mm
inch
(1)
Min
630
24.8
700
27.6 Foot end of rail to center line of WS detector.
With manual WS, the Min value should be larger than 630
mm (530+100 Cal space)
(2)
Min
203
8.0 400
15.7
Service
clearance for WS installation.
The Min value should be larger than 203 mm, the WS back
cover could be installed or can be removed.
(3)
Fix
1045
41.1
1045
41.1 Half of 2099mm Bridge.
(4)
Min
102
4.01
200
7.9 Room wall to drape rail
Make sure cable not touch the wall
(5)
Min
62
2.44
200
7.9 Drape rail to back of bridge.
Make sure cable could bend freely
(6)
Fix
2099
82.6 2099
82.6
2099mm
Bridge.
(7)
Min
500
19.7
1000
39.4 Service clearance for bridge installation.
(8)
Min
900
35.4
1000
39.4 Distance between head end of bridge and OTS when OTS at
head limit position.
(9)
Min
1900
74.8
2000
78.7 Travel range of OTS.
The Min value should be larger than 1900 mm(1800
SID+100 cal).