Section 3.0 - Room Layout
could be exposed. Remove all table cover. most components could be repaired except replacing table actuator
and tabletop.
Room Size Summary
1.) If with control room, width need add min 1100mm. so min room size is 4217mm*4800mm
2.) If 2200mm stretcher table is used, width need min 4000mm (200+200+960+table width/2+2200) to push it into
exposure room, so min room size is 4217mm*4000mm.
3.) Applicable to below 2 configurations:
Standard table+ manual WS
Standard table +G2 Std. WS
Clinical Impact when Chogori Std. Table+WS in Proteus Room with Min Ceiling Height 2500mm
If Chogori Std. Table+WS are installed in 2.5m ceiling height room, the clinical impact is:
1.) For table mode, the tabletop max height is 67.3cm while maintaining 100cm SID. This impact the operator to
bend for operation, it is manageable from the operator’s perspective. We have proven use of a similar height
in the Brivo 385 DRF which is a fixed height table. It is fixed at 70.1 cm. We think the difference of 2.8cm or
about 1 in. is small enough to equate the usability of this height.
2.) The max SID is 114.8cm when tabletop at lowest position, we are within a customer’s expectation concerning
the maximum available SID.
3.) For WS coverage, tube max height is 159.9cm, for 185cm (95% coverage) height patient, it can provide
coverage for great majority of chest exams, it’s usable. While many neck or head exams at the Wallstand may
require the patient to be seated.