GE Healthcare
Direction 5743854-1EN, Revision 1 Optima XR642/XR648 Pre-Installation
Page 84
Section 3.0 - Room Layout
(4)
Min
960
37.8
1000
39.4 Back of bridge to center line of table.
With manual WS, the Min value should be larger than 1960 (860
+100 Cal space + 600 distance between focal spot to back of OTS
400 half table Width)
(5)
Min
1100
43.3
1300
51.2 Table center line to WS center line.
The Min value should be larger than 1000 mm (700 distance
between focal spot to back of OTS 400 half table width)
(6)
Min
102
4.01
200
7.9 Room wall to drape rail
Make sure cable not touch the wall
(7)
Min
62
2.44
200
7.9 Drape rail to back of bridge.
Make sure cable could bend freely
(8)
Fix
3077
121.1
3077
121.1 3M Bridge length
(9)
Min
100
3.94
500
19.7 Service
clearance for OTS installation.
(10)
Min
50
2.0 100
3.9
Service
clearance
for Tabletop at head limit postion.
The min value for tabletop not touch the wall.
(11)
Fix
1660
65.4
1660
65.4 Half length of tabletop longitudinal travel coverage
(12)
Fix
660
26.0
660
26.0 Half length of table base (no tabletop)
(13)
Fix
2330
91.7 2330
91.7
Length
of
tabletop
(14)
Min
50
2.0
500
19.7 Service clearance for Tabletop pulled out.
The min value for tabletop not touch the wall when it is pulled
out. FE could stand by side of the tabletop if room length is short.
(15)
Min
1353
53.3
1400
55.1 Service clearance for cabinet.
(16)
Min
1307
51.5
1400
55.1 Service clearance for cabinet.
(17)
Min
200
7.9 200
7.9
Service clearance for cabinet.
Pre-Install.book Page 84 Tuesday, January 30, 2018 2:36 PM