Maintenance
3.
Connect the cables and harnesses, identified in the following photograph,
to the PWA.
4.
Position the power supply/KISS pump base on the chassis.
5.
Secure the power supply/KISS pump base to the chassis using its fasteners, as
identified in the following photograph.
6.
If it was installed, replace the KISS pump and tubing.
See step
through step
in
“Replacing the KISS Pump Assembly” on page 127
for instructions.
7.
Replace the power supply assembly.
See step
through step
in
“Replacing the Power Supply Assembly” on page
for instructions.
8.
Reattach the top cover assembly.
See
“Reattaching the Top Cover Assembly to the Device” on page 81
for
instructions.
9.
Reinsert the battery.
See step
through step
in
“Replacing the Battery” on page 74
for instructions.
10.
Continue with the appropriate functional checkout procedure for this FRU.
See
“Functional Checkout” on page 129
for more information.
88
MAC™ 2000 ECG Analysis System
2053535-003C