Electromagnetic Compatibility (EMC)
EMC Immunity Test
Immunity Test
Compliance
Test Level
Compliance
Level
Electromagnetic
Environment–Guidance
Do not use portable or mobile
RF communications equipment
closer to any part of the system,
including the cables, than
the recommended separation
distance calculated for the
equation applicable to the
frequency of the transmitter.
Conducted RF
3 Vrms
150 KHz to 80
MHz
@ 2 Hz mod.
Multilink Lead
set: 3.0 Vrms
Value Lead set:
3.0 Vrms
KISS Lead set:
0.3 Vrms
Recommended separation
distance:
Multilink Lead Set:
d
= 1.2√P
Value Lead Set:
d
= 1.2√P
KISS Lead Set:
d
= 11.6√P
168
MAC™ 2000 ECG Analysis System
2053535-003C