Maintenance
Replacing the Display Assembly
Use the following procedure to replace the display assembly.
1.
Insert the new display module into the display bezel.
2.
Replace and tighten the screws securing the display to the display holder, as
shown in the following photograph.
3.
Reconnect the display cable to the LCD display and affix the cable to the back of
the display, as shown in the following photograph.
NOTE:
The ferrite on the display cable has double-sided tape that allows it to be
gently affixed to the back of the display.
4.
Replace the display bottom cover, carefully threading the keypad ribbon and
display cable through the openings, as shown in the following photograph.
5.
Secure the display bottom cover with the ten screws, as shown in the following
photograph.
2053535-003C
MAC™ 2000 ECG Analysis System
111
Содержание MAC 2000
Страница 41: ...Product Overview 2053535 003C MAC 2000 ECG Analysis System 41 ...
Страница 134: ...Maintenance 134 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 160: ...Parts List 160 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 182: ...182 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 183: ......