Service Setup
5.
When the test is done, press
Esc
or
Cancel
.
The following window opens:
6.
Do one of the following, depending on the results of the test:
• If the test passed, press
Yes
• If the test failed, press
No
.
Replace the main PWA as described in
Testing the Patient Leadwires
Use the following procedure to test the patient leadwires:
1.
Open the
Diagnostic Tests
window as described in
page 46
2.
Connect a patient cable with leadwires to the device’s patient cable connector.
3.
Connect all leads to a patient simulator or shorting bar.
4.
Select
Patient Lead Wire Check
.
The following window opens:
5.
Press
Enter
to select
Start Test
.
The test results are displayed for each leadwire.
• If the following message is displayed, the leadwire passed the test:
Connected
.
• If the following message is displayed, the leadwire failed the test:
Disconnected
.
6.
When the test is done, press
Esc
or
Cancel
.
2053535-003C
MAC™ 2000 ECG Analysis System