Service Setup
In the case of wireless (WiFi) LAN, ensure that the wireless LAN settings are
configured as explained in Communication Setup in the
MAC™ 2000 ECG
Analysis System Operator’s Manual
.
2.
Open the
Diagnostic Tests
window as described in
page 46
3.
Select
LAN Test
.
The following window opens:
4.
Press
Enter
to start
Test network connectivity
.
The following message is displayed:
Checking connectivity. Please wait.
Then the test results are displayed.
• If the following message is displayed in the window, the test passes:
System
Connected to Network
.
• If the following message is displayed in the window, and you are sure the
system is connected to an active network, the test fails:
Network Unavailable
.
5.
When the test is done, press
Esc
or
Cancel
to close the results window.
If the test failed, replace the main PWA as described in
.
Testing the Internal Modem
Use the
Modem Test
to test the internal modem.
1.
Connect the device to an active analog phone line.
Ensure that the phone line is active. If you connect to an inactive phone line, the
test result may be a false negative.
2.
Open the
Diagnostic Tests
window as described in
page 46
3.
Select
Modem Test
.
The following window opens:
4.
Select
Internal Modem Test
, press
Enter
.
The following message is displayed in the window:
Test in Progress. Please
wait.
2053535-003C
MAC™ 2000 ECG Analysis System
57
Содержание MAC 2000
Страница 41: ...Product Overview 2053535 003C MAC 2000 ECG Analysis System 41 ...
Страница 134: ...Maintenance 134 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 160: ...Parts List 160 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 182: ...182 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 183: ......