Technical Specifications
Display (cont'd.)
Item
Specifications
Data
Heart rate, patient ID, clock, battery power indicator, waveforms,
lead labels, speed, gain and filter settings, warning messages,
prompts, and help messages.
Standard display: 6 leads
Optional display: 12 leads
Traces
3, 6, 12
Sweep Speed
12.5, 25, and 50 mm/s
Filtered ECG
The raw or filtered ECG curves are displayed on the screen as
selected by the user.
Keypad
Item
Specifications
Pad
Membrane board with tactile feedback
Keys
Soft function keys
Full alphanumeric keyboard (Qwerty key set)
Writer controls
Trimpad cursor controls
Keypad
Standard keypad with additional function keys.
Keypad and lettering are water resistant and meet MDM4021
Cleaning and Chemical Resistance Performance Guidelines
Functions
Standard
Stress
Printer
Item
Specifications
Technology
Integrated thermal dot array
Paper Speed
5, 12.5 mm/s at ±10% accuracy
25, 50 mm/s at ±2% accuracy
Printer motor speed
Report printing within 25 seconds
Number of traces
Up to 12 ECG traces sampled at 500 or 1000 Hz
Sensitivity/gain
2.5, 5, 10, 20, 40 mm/mV with ±5% accuracy
Speed accuracy
5, 12.5 mm/s at ±5%
25, 50 mm/s at ±2%
Amplitude accuracy
±5%
Horizontal resolution
40 dots/mm at 25 mm/s in rhythm mode
Vertical resolution
8 dots/mm
Paper type
Z-fold thermal with pre-printed grid and perforation
Paper size
215 mm x 280 mm (8.5 in x 11 in) (letter)
210 mm x 295 mm (8.25 in x 11.6 in) (A4)
214.2 mm x 279.4 mm (8.43 in x 11 in) (modified letter)
Paper tray capacity
Holds up to 150 sheets
172
MAC™ 2000 ECG Analysis System
2053535-003C
Содержание MAC 2000
Страница 41: ...Product Overview 2053535 003C MAC 2000 ECG Analysis System 41 ...
Страница 134: ...Maintenance 134 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 160: ...Parts List 160 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 182: ...182 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 183: ......