Maintenance
4.
Turn over the top cover assembly, slide the display assembly toward the front of
the top cover assembly, and lift the display assembly, as shown in the following
photograph.
NOTE:
Carefully guide the display cable and keypad ribbon out of the slots on the
top cover.
5.
Pry out the O-ring from the top cover, as shown in the following photograph.
6.
Turn the display over so the bottom display cover is facing you.
7.
Remove the ten screws from the display bottom cover, as identified in the
following photograph.
114
MAC™ 2000 ECG Analysis System
2053535-003C