Service Setup
After the device’s flash memory is formatted, you need to reconfigure the device,
Refer to the
MAC™ 2000 ECG Analysis System Operator’s Manual
for information on
configuring the device.
If formatting the device’s memory flash does not resolve the memory or data
corruption issues, you need to replace the main PWA. For instructions on replacing the
PWA, see
“Replacing the PWA” on page 124
.
Opening a Command Prompt
Use the following procedures to open a command prompt from which you can issue
various network diagnostic commands.
1.
Go to the
Service Setup
menu, as described in
page 43
2.
Select the
Open Command Prompt
button.
The following command prompt window opens:
3.
Issue any of the following commands as necessary.
Command
Description
ping<ipaddress>
The system queries and receives replies for the IP address
identified in the command.
ipconfig
Returns information related to the TCP/IP configuration of
the system.
netstat
Returns network statistics for the system.
netstat -e
Returns ethernet statistics for the network.
netstat -n
Returns system network addresses and port numbers in
numerical form.
netstat -p
Returns statistics specified for the IP4 protocol.
netstat —r
Returns the routing table.
4.
When you are done, type
exit
and press
Enter
to close the command prompt
window and return to the
Service Menu
.
70
MAC™ 2000 ECG Analysis System
2053535-003C
Содержание MAC 2000