4
Maintenance
This chapter addresses maintenance recommendations, required tools and supplies,
FRU replacement procedures, and functional checkout procedures.
Recommended Maintenance
Regular maintenance, irrespective of usage, is essential to ensure that the equipment
is always functional when required.
Refer to the operator’s manual for your device for cleaning procedures. In addition,
perform the printhead cleaning procedure, as described in
on page 99
. The system does not require any calibration procedures. GE Healthcare
recommends that you perform electrical safety checks annually. For more information,
see
“Electrical Safety Checks” on page 132
WARNING:
EQUIPMENT FAILURE AND HEALTH HAZARDS — Failure on the part of all responsible
individuals, hospitals or institutions, employing the use of this system, to
implement the recommended maintenance schedule, may cause equipment
failure and possible health hazards.
The sole responsibility rests with the individuals, hospitals, or institutions utilizing
the system.
The manufacturer does not, in any manner, assume the responsibility for
performing the recommended maintenance schedule, unless an Equipment
Maintenance Agreement exists.
Required Tools and Supplies
The following tools are required to perform the procedures described in this chapter:
• ECG simulator
• Small flat blade screw driver
• Posidriver PZ1 x 80
• Posidriver PZ2 x 100
• Current leakage tester
• Anti-static wrist strap
• Small side cutter pliers
2053535-003C
MAC™ 2000 ECG Analysis System
71
Содержание MAC 2000
Страница 41: ...Product Overview 2053535 003C MAC 2000 ECG Analysis System 41 ...
Страница 134: ...Maintenance 134 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 160: ...Parts List 160 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 182: ...182 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 183: ......