Maintenance
5.
Open the printer door and locate the latches at the top left and top right corners
of the printer compartment, just below the top cover, as shown in the following
photograph.
6.
Using a flat-head screwdriver, gently disengage the top cover from the latches.
7.
Gently lift the top cover part way up and detach the display cable (1) and keypad
ribbon (2).
8.
Lift the top cover off completely.
The top cover assembly, with the display assembly mounted on it, is now
completely separated from the device, as shown in the following photograph.
9.
Do one of the following:
• To service one of the device’s internal components, refer to the appropriate
FRU replacement procedure.
• To replace the top cover, proceed to
“Separating the Top Cover from the Top
, next.
76
MAC™ 2000 ECG Analysis System
2053535-003C