Maintenance
4.
Snap the back panel into place in the bottom cover.
5.
Reconnect all ribbons and cables to the PWA.
6.
Replace the top cover assembly.
See
“Replacing the Top Cover” on page 79
for instructions.
7.
Reinsert the battery.
See step
through step
in
“Replacing the Battery” on page 74
for instructions.
8.
Turn on the system.
9.
Use the following procedure to restore system setups that were saved to the SD
card before the new PWA was installed:
a.
Insert the SD card with the saved system setups in to the SD card slot.
b.
On the
Main Menu
, press
System Configuration
.
c.
Press
More
>
More
>
Import Setup
.
d.
Highlight the system setup file you want to import from the SD card.
e.
Press
Import
.
f.
When the following message is displayed, press
OK
:
Configuration was
successfully imported
g.
Remove the SD card and store it in a secure location.
10.
Configure the
Device Settings
:
a.
On the
Main Menu
, press
System Configuration
.
b.
Press
More
>
More
>
Service Setup
.
c.
When prompted, enter the
Service Password
.
If you do not know the password, contact GE Healthcare Technical
Support.
2053535-003C
MAC™ 2000 ECG Analysis System
125