Maintenance
3.
Move the focus to
Complete Setup
and press
Enter
.
The
System Setup
report prints.
4.
Save the printed setup report in a secure location.
You can use the report as a reference if you need to restore the system setup
manually.
NOTE:
Option codes are not part of system configuration settings. These codes
are listed on the last page of your printed setup report. These codes are
also found on a label next to the battery compartment. The printed list of
option codes is a more convenient aid to use if you need to re-enter them
after installing the new PWA.
Removing the PWA
Use the following procedure to remove the PWA.
1.
Before beginning any steps, follow the instructions in
FRU Replacement” on page 73
2.
Remove the top cover assembly.
See
“Removing the Top Cover Assembly” on page 75
for instructions.
3.
Disconnect all ribbons and cables from the PWA.
4.
Remove the rear connector panel.
122
MAC™ 2000 ECG Analysis System
2053535-003C