Service Setup
7.
Replace every leadwire that failed the test.
8.
Repeat the test.
If the leadwire still fails the test, replace the main PWA as described in
on page 120
.
Printing a Service Report
Use the following procedure to print a service report after completing any of the
diagnostic tests.
1.
Open the
DIAGNOSTIC TESTS
window as described in
on page 46
.
2.
Select
Service Report
and press
Enter
.
Updating the Software
Software updates are provided on an SD card. Perform a software update as
described in this section.
NOTE:
The AC power must be connected while you are updating the software through
the SD card.
Accessing the Software Update
1.
Insert the SD card with the software update in the SD card slot in the rear panel
as shown in the following photograph.
The gold contacts are face-down.
2.
On the
Main Menu
, press
System Configuration
.
60