Maintenance
Removing the Display Assembly
Use the following procedure to remove the display assembly.
1.
Before beginning any steps, follow the instructions in
FRU Replacement” on page 73
2.
Remove the top cover assembly.
See
“Removing the Top Cover Assembly” on page 75
for instructions.
3.
Turn over the top cover assembly and remove the six fasteners shown in the
following photograph.
4.
Turn over the top cover assembly, slide the display assembly toward the front of
the top cover assembly, and lift the display assembly, as shown in the following
photograph.
NOTE:
Carefully guide the display cable and keypad ribbon out of the slots on the
top cover.
5.
Pry out the O-ring from the top cover, as shown in the following photograph.
6.
Turn the display over so the bottom display cover is facing you.
2053535-003C