Maintenance
8.
Remove the display bottom cover, carefully threading the display cable and
keypad ribbon through the openings.
9.
Disconnect the display cable and remove it from the LCD display.
NOTE:
The ferrite on the display cable is affixed to the display with a double-sided
tape which can be gently removed.
10.
Remove the four screws from the display holder, as shown in the following
photograph.
11.
Remove the display holder and gently lift the LCD out of the display bezel.
Replacing the Keyboard
Use the following procedure to replace the keypad assembly.
1.
Insert the LCD display module into the new display bezel with the keypad FRU.
2.
Replace and tighten the screws securing the display to the display holder, as
shown in the following photograph.