Maintenance
2.
Position the printer motor on the printer module assembly and secure with the
two fasteners, as shown in the following photograph.
3.
Position the printer motor cover over the printer motor and secure with its
fastener, as shown in the following photograph.
4.
Reattach the printer module assembly to the bottom cover.
See
“Replacing the Printer Module Assembly” on page 95
for instructions.
5.
Continue with the appropriate functional checkout procedure for this FRU.
See
“Functional Checkout” on page 129
for more information.
Paper Sensor
The following procedures describe how to remove and replace the paper sensors.
Removing the Paper Sensor
Use the following procedure to remove the paper sensor from the print module
assembly.
1.
Before beginning any steps, follow the instructions in
FRU Replacement” on page 73
2.
Remove the printer module assembly.
See
“Removing the Printer Module Assembly” on page 94
for instructions.
2053535-003C
MAC™ 2000 ECG Analysis System
97