Maintenance
3.
Detach the KISS pump harness, identified in the following photograph, from
the PWA.
4.
Remove the fasteners securing the KISS pump assembly to its base.
5.
Lift the KISS pump assembly from its base.
Replacing the KISS Pump Assembly
Use the following procedure to replace the KISS pump assembly.
1.
Lower the KISS pump assembly into position on its base.
2.
Secure the KISS pump assembly to its base with its fasteners.
3.
Attach the KISS pump harness, identified in the following photograph, to the
PWA.
2053535-003C
MAC™ 2000 ECG Analysis System
127