Section 8-7 - Mechanical Option Installation instruction
8-7-8
Color Printer Fixture Middle installation - UP23MD Printer
1.) Install color printer on the top of the base bracket by screw 4 screws(2306562, FH M3x6 YEL).
2.) Unscrew 2 existing screws from the middle cover.
3.) Screw 1 screw (5176744, BH M4x20 WHT) on each side of the system to fix the Color printer with
base bracket to the system.
Figure 8-269 Color Printer Fixture Middle installation
Figure 8-270 Color Printer Fixture Middle installation