Section 8-4 - Software Installation
8-131
Section 8-4Software Installation
8-4-1
Installing Base Image Software
NOTE:
While we do not believe it is necessary, It would not hurt to disconnect the system from the
network and remove all transducers.
NOTE:
All digital peripherals, reporting systems connected via the USB ports should be either power
off or disconnected prior to the software installation
1.) Power ON the scanner.
2.) Tap gently the
DEL
key until the following screen appears.Enter “
arirang”
as a password if the
system request a password, then press the
Enter
key.
3.) Move to
Advanced
and select
IDE Configuration
.
Figure 8-179 Enter Current Password
Figure 8-180 IDE Configuration