GE H
EALTHCARE
D
IRECTION
5394141, R
EVISION
5
LOGIQ™ P5 S
EVICE
M
ANUAL
8-74
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-29
CL1TRX Assy / BL1TRX Assy
8-2-29-1
Tools
•
Common pilIips screwdrivers
8-2-29-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-29-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
8-2-16 "Side Right Cover" on page 8-52
.
2.) Remove the EMI Cover R. Refer to the
8-2-27 "EMI Cover R" on page 8-69
.
3.) Unscrew 2 screws (1-2) and eject the PCB.
4.) The CL1TRX(BL1TRX) board is located on the most left slot in the nest box.
Figure 8-109 Eject ASIG Assy