Section 5-10 - Service Platform
5-45
5-10-5
Diagnostics
NOTE:
When Service Login for FRU Diagnostic, scan mode must be B-mode.
Detailed
Diagnostic
information is found in Diagnostics/Troubleshooting.
NOTE:
After finishing Diagnostic test, the system should be rebooted.
5-10-5-1
Diagnostics Execution
Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level
diagnostic selection based on the user’s level and login access permissions. Remote access will require
disruptive diagnostic permissions to run Acquisition diagnostics.
5-10-5-2
Diagnostic Reports
Diagnostic tests return a report to the Service Platform. The platform retains the report and allows for
future viewing of the diagnostic logs.
5-10-5-3
Proactive Diagnostics
A system of self-monitoring is largely supported with the integration of
iLinq
. The scheduler, executive,
user interfaces, and some of the proactive diagnostic functions are provided by
iLinq
. Other tasks
should be provided by the product team.
WARNING
WARNING
Diagnostics should be performed with probes which have 128 or more elements.
Figure 5-43 GE Service Diagnostic Page