GE H
EALTHCARE
D
IRECTION
5394141, R
EVISION
5
LOGIQ™ P5 S
EVICE
M
ANUAL
Section 7-5 - Common Diagnostics
7-17
7-5-4
FRU Test
7-5-4-1
SYSCONPM/SYSCONCM
These programs are provided for testing the SYSCONPM/SYSCONCM board.
Table 7-2
FRU Test, SYSCONPM/SYSCONCM Diagnostics Menu
Menu
Descriptions
User
Level
Tools
Full Test
Performs all of the L1 SYSCON board tests listed below
All
None
Assy Revision Test
Test EEPROM.
Send test pattern to write to EEPROM and read
Displays the board revision of the SYSCONPM/SYSCONCM board
All
None
USC I/F Test
The CL1TRX, P3RLY, optional ACWD boards are connected to the SYSCONPM/
SYSCONCM via the USC bus.
The DSP in SYSCONPM/SYSCONCM accesses the boards above to check if the
USC bus is connected properly to each board.
All
None
FEBUS Test
Checks if The DSP in SYSCONPM/SYSCONCM accessed the boards, CL1TRX/
BL1TRX and ACWD, via the FEBUS.
All
None
TRIG Test
Check the trigger max, min and ever interval
All
None
PCI I/F Test
Sends the command to the DSP to check the SDRAM in SYSCONPM/SYSCONCM All
None
DSP Memory Test
Sends the command to the DSP to check the internal DSP memory
All
None
Board Temperature
Read SYSCON Board Temperature.
All
None
CHACOM I/F Test
Sends the command to read / write data from DSP to CHACOM
GE FE
Only
None