Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-9
8-2-3
OSD button and lamp set, LCD Plastic Filter
8-2-3-1
Tools
•
Common pilIips screwdrivers
8-2-3-2
Preparations
•
Shut down the system and switch off the main breaker.
•
Maneuver control console to a suitable position for removing the monitor.
8-2-3-3
Removal procedure
1.) Unscrew 4 screws (1-4) from the LCD front cover.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-3-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-12 Unscrews 4 screws
Table 8-4
Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394141, Rev 1+, Section 8-2-3. Equipment
passed all required tests and is ready for use. “
15" OSD Key Assy Function Validation
(1)
(2)
(3)
(4)