Section 7-7 - Trouble Shooting Tree
7-29
7-7-5
Trackball
Figure 7-16 Trackball Troubleshooting
The trackball is low sensitive!
Remove the ball from the Trackball,
and clean it.
YES
NO
Does the trackball move at
all?
Replace the trackball unit.