6.) Unscrew the 4 screws on the front of the DVD drive and remove the DVD Assy.
7.) Unscrew 8 screws to remove the bracket from the DVD Assy.
8.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-44-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-135 Unscrew 4screws and remove the DVD Assy
Figure 8-136 Unscrew 8screws and remove the DVD Drive from the bracket
Table 8-45 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394141, Rev 1+, Section 8-2-44. Equipment
passed all required tests and is ready for use. “