Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-8
Sub Keyboard Assy
8-2-8-1
Tools
•
Common pilIips screwdrivers
8-2-8-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-8-3
Removal procedure
1.) Remove the Main Keyboard Assy. Refer to the
8-2-7 "Main Keyboard Assy" on page 8-37
.
2.) Unscrew 4 screws (1-4). 2 screws(1-2) are back side or the speaker and 2 screws(3-4) are bottom
of sub keyboard.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Figure 8-70 4 Screw Sub Keyboard Assy
Table 8-9
Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394141, Rev 1+, Section 8-2-8. Equipment
passed all required tests and is ready for use. “
Sub Keyboard Assy Function Validation
(1)
(2)
(3)
(4)