GE H
EALTHCARE
D
IRECTION
5394141, R
EVISION
5
LOGIQ™ P5 S
EVICE
M
ANUAL
Section 8-4 - Software Installation
8-135
8-4-1 Installing Base Image Software
10.)Approx. twenty(20) minutes later, the following screen appears. Eject the Base Image Software
Disk from the CD/DVD ROM drive.
11.)After removing the disk from the drive, Press ALT + CTRL + DEL key.
NOTE:
After Installing Base Image Software, the “General failure reading drive A. Abort, Retry, Fail?”
message can be appeared, but Ignore it! All procedure would be successfully done.
12.)The following message box can be displayed after the SYSTEM boot up. Press Yes.
Figure 8-186 Installation Completion
WARNING
WARNING
Remove the disk from the drive. Otherwise you will be encountered the Base Image
Software installation screen iteratively.
Figure 8-187 System Setting Change