Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-9
Trackball Assy
8-2-9-1
Tools
•
Common pilIips screwdrivers
8-2-9-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-9-3
Removal procedure
1.) Remove the Main Keyboard Assy. Refer to the
8-2-7 "Main Keyboard Assy" on page 8-37
.
2.) Unscrew 4 screws (1-4) and disconnect the cable.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-9-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-73 Screws and connector for Trackball
Table 8-10 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394141, Rev 1+, Section 8-2-9. Equipment
passed all required tests and is ready for use. “
4 screws
connector