Section 8-6 - Software Reloading
8-6-2-1 Perform Software Reloading in the Service Platform
3.) Press “Time Zone”
4.) Register the current Time Zone
5.) Press “Cancel”.
6.) Log onto the Service Platform.
7.) Go to “Utilities”->”Common Utilities”.
8.) Go to “Utility”->”System”->”About”
Verify that Software version and software build version are same with the ones before the Software
Reloading.
Figure 8-232 Time Zone
Current Time Zone
Table 8-1 Current Time Zone