GE H
EALTHCARE
D
IRECTION
5394141, R
EVISION
5
LOGIQ™ P5 S
EVICE
M
ANUAL
Section 4-4 - Software Configuration Checks
4-25
4-3-12
ECG Check Out
Connect the ECG Harness and check:
Section 4-4
Software Configuration Checks
Table 4-5
ECG Harness Check
Step
Task to do
Expected Result(s)
1.
Connect the ECG at the Connector on the Front of the
scanner.
It will display a curve along the bottom edge of the image
sector
Table 4-6
Software Configuration Checks
Step
Task to do
Expected Result(s)
1.
Check Date and Time setting
Date and Time are correct
2.
Check that Location (Hospital Name) is correct
Location Name is correct
3.
Check Language settings
Desired Language is displayed
4.
Check assignment of Printer Keys and User Define Keys
Print1, 2 Keys and User define Keys are assigned as
desired by the customer
5.
Check that all of the customer’s options are set up correct All authorized functions are enabled
Содержание LOGIQ P5
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION 5394141 REVISION 5 LOGIQ P5 SEVICE MANUAL i 9 JA ZH CN KO...
Страница 28: ...GE HEALTHCARE DIRECTION 5394141 REVISION 5 LOGIQ P5 SERVICE MANUAL i 28...
Страница 531: ......