GE H
EALTHCARE
D
IRECTION
5394141, R
EVISION
5
LOGIQ™ P5 S
EVICE
M
ANUAL
7-22
Section 7-6 - LED Descriptions
7-6-4
APS/APS Pro Assy
NOTE:
HVL+ and HVL- are not available for APS PRO.
The APS/APS Pro Assy is located at the No. 4 slot of the NEST Assy.
Table 7-11
LEDs on APS/APS Pro Assy
LED
Descriptions
Normal State
DP5 (Top)
Digital 5V output indicator
Normally On
LP5
Live Power 5V output indicator
Normally On
AP6
Analog 6V output indicator
Normally On
DP33
Digital 3.3V output indicator
Normally On
SHV+
SHV+ output indicator status indicator
Normally On
HVL+
HVL+ level indicator
Normally On in CFM mode or PW mode
HVH+ (Bottom)
HVH+ level indicator
Normally On in B mode