Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-3
8-2-1
LCD Plastic filter
8-2-1-1
Tools
•
Common Phillips screwdrivers
•
Allen/Unbraco wrench
•
Stubby screwdriver (Flat tip and Cross tip)
8-2-1-2
Removal procedure
1.) Remove the LP5 Monitor cover set. Refer to the
8-2-2 "Monitor cover set" on page 8-5
2.) Remove the LCD Plastic filter.
3.) Put the LCD plastic filter on the LCD monitor after detaching the tape of the LCD plastic filter.
Figure 8-1 Removing the LCD plastic filter
Figure 8-2 Putting the LCD plastic filter on the LCD monitor