Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-32
ACWD Assy
8-2-32-1
Tools
•
Common pilIips screwdrivers
8-2-32-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-32-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
8-2-16 "Side Right Cover" on page 8-52
.
2.) Remove the EMI Cover R. Refer to the
8-2-27 "EMI Cover R" on page 8-69
.
3.) Remove the ASIG Board. Refer to the
8-2-35 "ASIG Assy" on page 8-86
.
4.) Pull the ACWD assy out from Nest box. The ACWD assy is located in slot #2.
Figure 8-115 PWAs in Nest Box