A:64
AK 98 Dialysis Machine - Handling the dialysis machine
HCEN12745 Revision 10.2016
Program version 2.xx
●
Stroke volume
: Read out of actual stroke volume.
●
Mean blood flow
: Read out of effective blood flow rate during treatment.
Tools
●
Service
: Open the
Service menu
.
●
Clean screen
: Allows the operator to clean the screen when the machine is
switched on. It is possible to perform this procedure in any machine status. When
the command has been activated the screen is temporarily disabled, allowing the
operator to clean it.
●
Night light
: Turns off the operator´s panel. Press the operator´s panel at any
time to reactivate it. Any new alarm reactivates the operator´s panel.
3.4.20
Fluid button (25)
The
Fluid
button opens the
Fluid Menu
, which contain tabs for handling parameters
related to the fluid path.
WARNING!
The maximum UF rate must be preset by an authorised technician according to the
patient’s prescription, when AK 98 dialysis machine is used in home healthcare
environment.
UF rate
●
UF volume
: Set the ultrafiltration volume (patient weight loss).
●
Treatment time
: Adjust the treatment time. Time menu will display total set
treatment time.
●
Set minimum UF rate
: Set the minimum ultrafiltration rate to be applied, as
required during the treatment.
●
: Set the isolated ultrafiltration values, see Section 6
●
Profiling
: Set the ultrafiltration profile, see Section 7.1
on page A:118.
●
Calculated UF rate
: Read out of the calculated ultrafiltration rate.
●
Accumulated UF volume
: Read out of accumulated ultrafiltration volume (patient
weight loss).
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...