AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
201 Blood flow is too low
201 Blood flow is too low
Appears:
When the set blood flow rate has been lower than, or equal
to, 100 mL/min for more than 5 minutes.
Machine actions:
None.
Operator action:
1.
Check for a cause, for example: recurring problems
with the needle, previous alarms that took a long time to
solve and technical alarms.
2.
Press the
Blood pump
button to start the blood pump
when the cause of the alarm has been corrected.
3.
If possible, increase the blood flow rate to the intended
value.
4.
If necessary, adjust the low alarm limit.
- Press the
Blood
button.
- Select the
Blood flow
tab.
Press
Low alarm limit
and set a suitable alarm limit
for low blood flow.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...