B:112
AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
538 Fluid tube sensor test
538 Fluid tube sensor test
Attach fluid tubes to safety couplings. To
interrupt test press Confirm.
Appears:
When the dialysis machine cannot detect that the dialysis
fluid tubes are attached to the safety couplings. The
corresponding machine test cannot be made.
Machine actions:
The ongoing disinfection or rinse program stops.
Operator action:
To continue the disinfection or rinse program:
1.
Check that the dialysis fluid tubes are properly attached
to the safety couplings.
2.
Press
Confirm
. The disinfection or rinse program
continues.
As an alternative, consult an authorised technician.
Mute time:
2 minutes.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...