B:16
AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
4.
Simultaneously, raise the blood level in the drip
chamber with the level adjustment knob by turning the
knob anticlockwise.
5.
When the blood level is correct and the alarm text has
changed, check that there is no air in the venous and
arterial line.
6.
Check that there is no air in the infusion line.
7.
Start the blood pump by pressing
Confirm
in the alarm
message.
8.
Adjust the blood flow rate.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...