A:112
AK 98 Dialysis Machine - Haemodialysis - Single needle treatment
HCEN12745 Revision 10.2016
Program version 2.xx
3)
Press the flashing
Ultrafiltration
button to start ultrafiltration.
4)
Press the flashing
Arterial
pressure
control to centralise the alarm limits around
the actual value.
5)
Press the flashing
Venous
pressure control
to set the venous pressure alarm limits.
Alarm limits for venous pressure are
automatically set to +350 mmHg (high
limit) and +150 mmHg (low limit).
6)
If needed adjust the venous pressure high
alarm limit using the arrow buttons to set
the desired value.
7)
Gradually adjust the venous pressure low
alarm limit during the arterial phase, where
blood circulates from the patient to the
dialysis machine.
8)
If necessary, adjust the blood level in the
expansion chamber.
5.5
End a single needle treatment
When the treatment is finished and treatment time on the screen has counted down to
0:00, there will be an attention.
596 Treatment time expired
To discontinue treatment press Confirm.
Procedure
1)
Press
Confirm
to confirm treatment time
expired.
2)
Press the
Blood pump
button to stop the
blood pump.
3)
Select the
Single needle
tab.
4)
Deactivate single needle treatment. Close
the menu.
5)
End the treatment. Follow the steps in
Section 4.3.1
on
page A:100.
6)
Remove the dialyzer and the blood lines.
Follow the steps in Section 4.3.3
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...