HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Technical data and specifications
A:183
Aproximate time
Rinse/Drain
10 minutes
Aproximate time Drain 4 minutes
14.1.15
Disinfection and cleaning – exterior cleaning
All outside parts of the machine can be cleaned with ethanol (70 %) or isopropanol (60
%).
14.1.16
Water supply
WARNING!
To ensure the quality of the dialysis fluid, inlet water quality shall comply with
appropriate regulations and as a minimum requirement according to ISO 13959.
The table below is presenting the requirements for the water supply required for the
AK 98 dialysis machine.
AK 98 dialysis
machine water
consumption
During treatment: maximum 770 mL/min
During disinfection: 800 mL/min with peaks of 1100 mL/min not
exceeding 40 seconds.
Minimum inlet
pressure
0.12 MPa (1.2 bar)
Maximum inlet
pressure
0.6 MPa (6 bar)
Inlet temperature
Treatment: +5 to +30 ºC
Disinfection: +5 to +90 ºC
Connector in/outlet
Diameter 8 mm
Quality
Inlet water quality shall comply with appropriate regulations and as a
minimum requirement according to ISO 13959.
Level for conductivity shall not exceed 0.1mS/cm. It is possible to use
water with higher conductivity if it consists mainly of sodium salts. This
may however affect the accuracy of the fluid composition.
NOTE!
Local regulations may require the use of separation devices
in the supply and special measures to protect against the
possibility of back-syphonage from dialysis equipment into
the water supply.
Drain
The drain tube outlet shall be placed between floor level and maximum
1.2 m above the outlet connection from the fluid monitor. An air gap to
atmospheric pressure must always be arranged at the tube outlet.
Length of drain tube
≤10 m
14.1.17
Power supply
Mains voltage
115 V AC, 50 Hz
115 V AC, 60 Hz
230 V AC, 50 Hz
230 V AC, 60 Hz
Protection class
Machine: Class 1 type B
BPM: Type BF
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...